1 Vmed Technology Vet-Dop2 TM Doppler Blood Pressure System Model BF2 User Manual
10 Doppler Procedure Doppler detects blood flow in arteries and veins. The radial, cranial, tibial, and median coccygeal arteries are
11 5. Select an animal cuff with a width approximately equal to the width of the limb and up to two times the width of the forearm mea
12 Tips for using the Doppler: Review the original Vet-Dop procedure video. (NOTE: This video shows the older Vet-Dop, but the procedure is the sam
13 Clinical References Hypertension Cats: Systolic pressure greater than 170-180 mmHg Dogs: Systolic pressure greater than 180 mmHg Hypertens
14 Chapter 3 Service, Maintenance and Specifications Service and Support Phone: 800-926-9622 FAX: 425-585-0231 Email: [email protected]
15 Maintenance There are no field serviceable components inside the Vet-Dop2. Opening the housing voids the warranty. Store this device in the st
16 Troubleshooting A faulty sensor or defective electronic components or both can cause excess noise, static, and difficulty acquiring a signal.
17 Appendix A Limited Warranty All Vmed Technology products carry the limited warranties described below. All warranty claims must be received by
18 Exclusive Remedies of Customer; Limitation of Liability. Vmed Technology’s liability and Customer’s remedy for damages incurred as
19 Appendix B Accessories and Replacement Parts Doppler sensor, 9.4 MHZ Part BF-94A (cats, dogs and exotics) BF-94R rats Certified Sphygmom
2 Table of Contents ACCESSORIES AND COMPONENTS ... 3 PRECA
20 Vet-Dop2 Doppler Blood Pressure System Model BF2 Vmed Technology, 2011 914 164th Street SE Mill Creek, WA 98012 Phone 800-926-9622 Fax 425
3 ACCESSORIES AND COMPONENTS Check Vet-Dop2 storage pouch and ensure the following items are included Vet-Dop2 Doppler Sphygmomanometer He
4 Chapter 1 Description Vet-Dop2 Doppler Vmed designed the Vet-Dop2 for veterinarians to measure systolic blood pressure on awake animals; to
5 Doppler Technique Doppler detects blood flow in arteries and veins. Doppler blood pressure is a manual technique using a Doppler flow detecto
6 Control Panel Figure 1 – Top panel with controls and indicator lights YELLOW LED here YELLOW or GREEN LED here RED LED he
7 Connector Panel The connector panel includes the battery charge port and the Doppler sensor receptacle as shown below. Figure 2
8 Chapter 2 Preparation and Use Doppler Preparation Ensure that the Vet-Dop2 is fully charged and assemble the unit, sensor, headset, appropriate
9 Cuff Selection Choose the appropriate cuff as shown in Figures 4 and 5. Figure 4 – Reference mark should fall within range Figure 5 – Referen